Clinical Investigation to Evaluate the CONVIVO System for Discrimination of Normal From Abnormal Tissue During Brain Tumor Resection
Summary
The primary objective of this study is to evaluate the diagnostic performance of the CONVIVO confocal endomicroscope in discriminating between normal and abnormal tissue in vivo during brain tumor surgery. The interpretation of intraoperative images obtained in situ will be tested against conventional histologic evaluation of targeted biopsies from imaged tissue. The study team hypothesize that there will be a high degree of correlation between images obtained with the CONVIVO system and conventional histologic interpretation.
Detailed description
The surgical resection of brain tumors is an integral component of modern neuro-oncology. . Extent of resection has been found to be associated with increased overall and progression free survival, with the greatest benefit occurring in the setting of a complete or gross total resection. Unfortunately, there are a number of obstacles unique to brain tumors that may limit the extent of resection. A number of advancements in neurosurgical oncology have emerged to improve the extent of tumor resection while decreasing operative morbidity and mortality. Confocal reflectance microscopy is a routine technique used to visualize tissues without fixation or staining used in classical histological techniques. Laser scanning confocal microscopy (LSCM) is an optical fluorescence imaging modality used for imaging thick in vivo and ex vivo tissues. Clinically available confocal endomicroscopy systems contain lasers with precise excitation wavelengths and dichroic filters for detecting appropriate emission wave lengths. The studies completed to date do not evaluate the ability of confocal microscopy to discriminate between normal and abnormal tissue at the margins during surgical resection. This is the first of such in vivo feasibility studies that aims to demonstrate this claim of the CONVIVO system. The images acquired with the CONVIVO system following administration of fluorescein will be compared to conventional histologic specimens from corresponding biopsies. This is a planned single center study. This study is designed to assess the diagnostic accuracy of the CONVIVO system compared with gold-standard histopathology in tissue that has already been identified for resection. The device will not be used to inform surgical decision making, nor will tissue that would not otherwise be resected be biopsied for research purposes. This study will rely on study investigators, all physicians at Dartmouth-Hitchcock, to identify subjects. Written consent is required for participation in the study.. There are no specific medical risks to patients associated with the use of CONVIVO. It is not anticipated that there are immediate or direct benefits to patients participating in this study. The CONVIVO imaging data acquired during the procedure will be initially stored on the device's hard drive and later securely uploaded to an encrypted Dartmouth-Hitchcock server. Patients will have the ability to maintain their privacy with minimal disruption or contact by the study team. Additionally, the PI or other investigators will have interactions with the patient as part of their routine clinical care. Participants are free to withdraw from participation in the study at any time upon request.
Arms & interventions
- Diagnostic TestCONVIVO system
Approximately 2-5 minutes following administration of FNa in situ imaging will be performed by the participating surgeon ensuring proper technique. Prior to entering the surgical field, the probe will be covered in a disposable sterile sheath that is manufactured with quality assurance for this purpose. The probe will gently be held against the tissue interface while imaging occurs. Again this will only be in regions that would normally be resected or sampled in routine clinical care. Following image acquisition, a neuropathologist present in the operating room, will review and capture each image.
- OtherConventional histologic evaluation
Following image acquisition, the tissue region imaged with the CONVIVO system will then be biopsied using biopsy forceps. This will be passed immediately off the surgical field as a research specimen and provided to a member of the research team to be prepared for conventional histologic evaluation. The specimen will be labeled with the deidentified subject and sample number. This sequence will then be repeated for each successive sample.
Outcome measures
Primary
Diagnostic performance of CONVIVO in discriminating between normal and abnormal tissue
Concordance between in vivo imaging with the CONVIVO system and conventional histologic assessment.The diagnostic performance of the CONVIVO system will be assessed with respect to classification as "normal" or "abnormal" tissue. The interpretation of intraoperative images obtained in situ will be tested against conventional histologic evaluation of targeted biopsies from imaged tissue. Pertinent tumors include all intracranial neoplasms including low and high grade glial neoplasms, glioneuronal tumors, cerebral metastasis, meningioma, schwannoma, and pituitary lesions.
Time frame: On day of Surgery, Day 0
Secondary
Degree of concordance with histologic diagnosis
Time frame: On day of surgery, Day 0
Sensitivity and Specificity of CONVIVO system
Time frame: On day of surgery, Day 0
Rate of uninterpretable or non-diagnostic imaging with the CONVIVO system.
Time frame: On day of surgery, Day 0
Correlation between intra-operative confocal microscopy and co-registered MRI points.
Time frame: On day of surgery, Day 0
Correlation between visible yellow-light fluorescence (560 nm filter) and intraoperative confocal microscopy.
Time frame: On day of surgery, Day 0
Correlation between visible blue-light fluorescence and quantitative fluorescence measurements
Time frame: On day of surgery, Day 0
Extent of tumor resection
Time frame: 2 Days after surgery up to one month.
Adverse events related to the administration of fluorescein
Time frame: Day of surgery (Day 0), 2 Days after surgery up to one month.
Time to interpretation of imaging and histology
Time frame: Day of surgery (Day 0), and up to one week
Ability to discriminate between viable tumor and pseudoprogression or "treatment effect" using the CONVIVO system.
Time frame: Day of surgery (Day 0) and up to one month
Eligibility criteria
Study locations (1)
Dartmouth-Hitchcock Medical Center